タイトル: Development of Repeatable Microcatheter Access Port for Intra-arterial Therapy of Liver Cancer.
その他のタイトル: 肝癌に対する経動脈的治療におけるマイクロカテーテル挿入可能ポー トの開発
著者: Fukuoka, Yasushi
Tanaka, Toshihiro
Nishiofuku, Hideyuki
Sato, Takeshi
Masada, Tetsuya
Tatsumoto, Shota
Marugami, Nagaaki
Sakaguchi, Hiroshi
Kichikawa, Kimihiko
キーワード: Transarterial chemoembolization
Hepatic arterial infusion chemotherapy
Implantable port
発行日: 2019年2月
出版者: Springer Nature
引用: Cardiovascular and interventional radiology Vol.42 No.2 p.298-303 (2019 Feb)
抄録: PURPOSE: To develop an implantable port in which a microcatheter can be inserted for a combination therapy of repeated transarterial chemoembolization (TACE) and hepatic arterial infusion chemotherapy (HAIC) for advanced liver cancer. MATERIALS AND METHODS: The design of a currently used implantable port was modified. A funnel part was constructed in the port. The septum was punctured by a 20-gauge indwelling needle, and 2.0-Fr non-tapered microcatheter was inserted into the port. In the in vitro studies, the advance of a microcatheter out of the funnel part was evaluated via seven different septum puncture sites. A 5-Fr indwelling catheter connected to the port was placed in a vascular model, and a microcatheter catheterization was evaluated. In an in vivo study, the port-catheter system was implanted in the hepatic artery in a pig. A microcatheter was percutaneously inserted through the port into the hepatic arterial branches, and embolization was performed. RESULTS: In the in vitro studies, the microcatheter was smoothly advanced out of the port and catheterizations into the hepatic arteries were successful via all septum puncture sites. In the in vivo study, repeated selective embolization through the port was successfully conducted on 7, 14 and 21 days after the implantation. CONCLUSION: The developed implantable port can be used for repeated catheter insertion into the hepatic artery. The combination of repeated TACE and HAIC could be possible using this device.
内容記述: 博士(医学)・甲第710号・令和元年6月26日
© Springer Science+Business Media, LLC, part of Springer Nature and the Cardiovascular and Interventional Radiological Society of Europe (CIRSE) 2018
This is a post-peer-review, pre-copyedit version of an article published in Cardiovascular and interventional radiology. The final authenticated version is available online at: http://dx.doi.org/10.1007/s00270-018-2111-0.
URI: http://hdl.handle.net/10564/3582
ISSN: 01741551
DOI: https://doi.org/10.1007/s00270-018-2111-0
学位授与番号: 24601A710
学位授与年月日: 2019-06-26
学位名: 博士(医学)
学位授与機関: 奈良県立医科大学