抄録: | Corneal configuration change due to recent keratorefractive surgery such
as photorefractive keratectomy (PRK), photherapeutic keratectomy (PTK), and laser in-
situ keratomileusis (LASIK) reduce the corneal rigidity, and will result in significant
underestimation of postoperative intraocular pressure (IOP) measurement. So we have
analyzed the relation between corneal cofiguration change and postoperative measurement
of IOP in keratorefractive surgery (PRK PTK, and LASIK).
Actual IOP was controlled using a manometer inserted into the anterior chamber of
enucleated pig eyes. Clinical IOP was measured with electronic applanation tonometer
(Tono-Pen XL, Menter) at several heights of manometer, central corneal thickness (CCT)
was measured with ultrasonic pachymeter (SP-2000, Tome.) before and after IOP measure-
ment, and anterior corneal curvature with Orbscan (Orbtech). Each measured IOP was
analyzed statistically with linear regression and paired t-test.
Measured IOP of the group of normal eyeball with various CCT was increased in every
experimentation as CCT increased. And measured IOP of automated lamellar keratoplasty
(ALK) performed group was significantly reduced as CCT reduced. Linear correlation was
established between set up IOP and measured IOP, with correlation coefficients close to
unity. Correlation between CCT and the measured IOP, CCT and the difference of the
measured IOP to each sample was statistically highly significant (P<0.001). Measured
anterior corneal curvature at 3 mm diameter of central cornea of the ALK performed group
was decreased as IOP increased.
Results of our investigation show that CCT has a larger effect on the measurement of IOP
than central corneal curvature and corneal hydration. We should take into consideration
the corneal configuration change for proper assessment of postoperative IOP measurement
in recent keratorefractive surgery. |